jagomart
digital resources
picture1_Light Microscopy Pdf 88643 | Els Brighdarkfield


 160x       Filetype PDF       File size 0.30 MB       Source: homepage.univie.ac.at


File: Light Microscopy Pdf 88643 | Els Brighdarkfield
light microscopy secondaryarticle brightfield and darkfield article contents brightfield kohler illumination theory illumination brightfield kohler illumination practice darkfieldillumination theory ronaldjoldfield macquarieuniversity sydney australia darkfieldillumination dryandimmersionsystems rheinbergillumination themicroscopeimage especiallyits contrast is ...

icon picture PDF Filetype PDF | Posted on 15 Sep 2022 | 3 years ago
Partial capture of text on file.
            Light Microscopy –                                                                               Secondaryarticle
            Brightfield and Darkfield                                                                            Article Contents
                                                                                                                 ¨
                                                                                                      . Brightfield(Kohler)Illumination – Theory
                                                                                                                 ¨
            Illumination                                                                              . Brightfield(Kohler)Illumination – Practice
                                                                                                      . DarkfieldIllumination – Theory
            RonaldJOldfield,MacquarieUniversity,Sydney,Australia                                      . DarkfieldIllumination – DryandImmersionSystems
                                                                                                      . RheinbergIllumination
            Themicroscopeimage,especiallyits contrast, is profoundly influenced by the way that       . Practical Applications of Darkfield Microscopy
            light is directed on to the specimen. The essential controls required are of intensity,   . Filters in Light Microscopy
            evenness, area of illumination and angle of illumination.
                                  ¨
            Brightfield (Kohler) Illumination –                               considerably to the cost of a microscope, but extends its
            Theory                                                            performance and efficiency.
                                                                                 An illumination system for professional microscopy
            Thecondenser                                                      mustsatisfy the following:
            The article ‘Light Microscopy’ describes the role of the          . The specimen field should be fully and evenly illumi-
            condenser and gives definitions of terms used in this                 nated.
            section. Withhistheoriesofmicroscopelensperformance,              . Theintensity of the illuminated field should be variable
            Ernst Abbe determined the requirements of specimen                   within wide limits.
            illumination. Essentially, the light source should comple-        . Thediameteroftheilluminatedfieldshouldbevariable
            tely fill the back focal plane (bfp) of the objective. This is        to match the fields of view of the more commonly used
            achieved by having a condenser for each objective,                   objectives, from about 0.2mm to 4mm.
            matching the objective numerical aperture (NA). The               . The angular aperture (NA) of the condenser should be
            microscope illuminator will include a lens or lenses that            variable to match the range of objective NAs, from
            may act as part of the condenser. This extends the                   about 0.1 to 1.3.
            complexity of the practical use of the condenser; manu-           . Selection of wavelengths (colour) may be desirable for
            facturers have devised many variations of condensers and             someaspectsofspecialized microscopy.
            illuminators, the choice of which depends on quality and
            price that can be afforded.
                                                                              Kohler illumination employs a collector lens, close to the
                                                                                 ¨
            Source-focusedillumination                                        filament, to form an image of the light source in the
                                                                              condenserlens‘aperture’,i.e.theplaneofthecondenseriris
            Earlier instruments focused an image of the light source          diaphragm (better called the aperture diaphragm). The
            itself, such as the broad flame of an oil lamp, on to the          filament image should completely fill the diameter of the
            specimen. Obviously, focusing the filament of a modern             aperture diaphragm. This enables an image of the light
            electric lamp on to the specimen will give very uneven            source to be formed in the bfp of the objective; the light
            lighting, and such a technique is not favoured today. In          source image will fill that plane, a requirement for the
            most microscopes of teaching laboratories, some modifi-            illuminationofthefullNAoftheobjective.Thefunctionof
            cationofsource-focusedilluminationisemployedbecause               the aperture iris is to change the illuminating NA, and to
            of its cheapness. At its simplest, a ground glass filter is        matchittotheobjective in use.
            interposed between the light source and the condenser to             Adjacenttothecollectorlensisthefield(iris)diaphragm.
            present a more evenly illuminated specimen.                       Thecondenserlensfocusesanimageofthefielddiaphragm
                                                                              ontotheplaneofthespecimen.Asthefieldirisis opened
                                                                              andclosed,thediameteroftheilluminatedspecimenfieldis
              ¨                                                               changed. To maximize contrast, no more of the specimen
            Kohlerillumination
                                                                              should be illuminated than is actually observed (or
            August Kohler (1866–1948) devised an illumination                 photographed).
                         ¨
            system specifically for photography through the micro-                All textbooks will have diagrams attempting to explain
                                                                                                     Figure1) but they are quite intimidat-
            scope. It is the system now acknowledged by all                   Kohlerillumination(
                                                                                 ¨
            manufacturers as being preferred for simple observation           ing until one understands that two different ray paths are
            as well as for recording microscope images. It adds               beingconsideredintheonefigure.Inaddition,eachofthe
                                                 ENCYCLOPEDIA OF LIFE SCIENCES © 2002, John Wiley & Sons, Ltd. www.els.net                 1
                 Light Microscopy – Brightfield and Darkfield Illumination
                                Retina                                               looking down a microscope? How can I superimpose an
                                                                                     image of a measuring device on the specimen?) may be
                                                                                     answeredwithreferencetoKohlerilluminationdiagrams.
                                                                                                                         ¨
                                                        Eyepiece                        Theseparate functions of the two iris diaphragms, field
                                              EYE      exit pupil                    and aperture, are not appreciated by many microscope
                                                                                     users, and is a frequent source of serious error in
                                                                                     microscope manipulation.
                                Intermediate
                                image      EYEPIECE
                                                                                                             ¨
                                                                                     Brightfield (Kohler) Illumination –
                                                                                     Practice
                                                                                     There is a complexity to the instructional diagrams and
                                                                                     written instructions for Kohler illumination. In addition,
                                                       Objective                                                     ¨
                                                 back focal plane                    eachbrandofmicroscope–eachmodel–islikelytohavea
                                                                                     different set of instructions. But there are three sequential
                                          OBJECTIVE                                  steps commontoall systems:
                                Specimen
                                                                                     1.   An image of the filament is focused on to the
                                                                                          (condenser) aperture diaphragm. In older micro-
                                         CONDENSOR                                        scopes, the lamp collector lens was focused to form a
                                                        Aperture
                                                  iris diaphragm                          sharp image of the filament on the aperture dia-
                                                                                          phragm. Modern systems are said to have their light
                                                                                          sources prefocused and precentred, and include
                                                                                          groundglass filters to make it impossible to check this
                                                                                          step.
                                                                                     2.   Animageofthefieldirisdiaphragmisfocusedonthe
                                                                                          specimen. The condenser is raised or lowered for this
                                                                                          step, and lamp-centring facilities will have to be
                                                                                          engaged. The field iris is then opened or closed until
                                                                                          its image matches the size of the observed specimen
                                                                                          field.
                                                                                     3.   The condenser NA is matched to the objective NA.
                                Field                                                     The objective bfp is examined (after removing an
                                iris diaphragm
                                            LAMP                                          eyepiece) to check the ‘seven-eighths’ position of the
                                         COLLECTOR                                        aperture iris (see the article ‘Light Microscopy’).
                                             LENS                                         Imperfect specimens may demand further adjustment
                                                                                          oftheapertureiristoenhancecontrastand/ordepthof
                                                        Filament                          field.
             Figure 1   Kohler illumination. The diagram on the left highlights the
                         ¨
             image-formingraypaththatincorporatesthefour conjugateplanes             Witheachchangeofobjective, the NA and field diameter
             associated with the specimen plane. On the right, the illuminating or   also change; strict Kohler illumination therefore requires
             aperture ray path is emphasized – four conjugate planes that incorporate                          ¨
             the filament and the lens ‘apertures’.                                  re-adjustment of steps 2 and 3.
                                                                                        AcondenserwithNAgreaterthan1.0isdesignedtobe
                                                                                     ‘immersed’ with a drop of immersion oil between the
             two ray paths has four conjugate planes, planes that are                condenser top lens and the slide undersurface. Although,
             images of each other, images on each other of preceding                 because it is messy, this is rarely done in routine
             planes. Familiarity with these two sets of four planes                  microscopy, it is quite appropriate for those instances
             greatly facilitates understanding of the microscope and is              when the very highest level of microscope performance is
             assumed knowledge in the instructions for more elegant                  sought.
             techniques such as differential interference contrast and                   Intensity control is by a variable transformer or
             fluorescence. Rather simple questions with quite difficult                 rheostat; the aperture iris should not be used for this
             answers (such as Where should the eye be placed when                    purpose.
             2
                                                                                       Light Microscopy – Brightfield and Darkfield Illumination
             DarkfieldIllumination – Theory                                          DarkfieldIllumination – Dry and
             It greatly limits the potential of the microscope to accept             ImmersionSystems
             that the role of the instrument is merely to produce an
             image that is an exact, enlarged copy of the specimen.                  The simplest darkfield attachment is a homemade patch-
             Especially with living material, contrast (i.e. variations in           stop(Figure3), a clear disc with an opaque centre. The disc
             theimageofcolourorintensity)maybeverypoor,andthe                        is inserted in the normal brightfield Abbe condenser, close
             imagepractically invisible. It is a common experience that              tothepositionofthecondenseraperturediaphragm;older
             smalldustparticlesintheairareeasiertodiscerniflightis                   microscopeshadafilterholderconvenientforthispurpose.
             directed not into the eye, but across the observer’s line of            Theaperturediaphragmitselfmustbeleftfullyopen.The
             vision.Althoughthedustistoosmalltobe‘resolved’bythe                     opaque centre of the disc prevents ‘direct’ illumination
             eye, particulate matter is seen, detected or made visible,              from entering the objective; the diameter of the opaque
             even to the extent that measurement could be made of                    region is not particularly critical, roughly one-half to two-
             particle movement. The same principle is used in micro-                 thirds the diameterofthecleardisc.Thetransparentouter
             scopybypreventingdirectilluminatingraysfromentering                     margins of the patchstop transmit a hollow cone of light
             theobjective. Darkfieldmicroscopy(manyprefertheterm                      towards, but missing, the objective. Only if a specimen is
             ‘darkground’) renders the object as bright against a dark               present will diffracted or reflected rays be redirected and
             background, considerably enhancing the contrast and                     accepted by the objective. In spite of its rather amateur
             visibility of small objects (Figure 2).                                 status,thepatchstopisaveryusefulmicroscopeaccessory,
             Figure 2   Plasmodiumvivaxinahumanbloodsmear;brightfield(left),darkfield (right). The bar represents 10 mm.
             Figure 3   Equipmentfordarkfieldmicroscopy.Fromtheleft:apatchstop,a drydarkfieldcondenser,animmersiondarkfieldcondenser,andanoil-
             immersionobjectivewithiris diaphragm.
                                                                                                                                                        3
                 Light Microscopy – Brightfield and Darkfield Illumination
             capable of serving objectives of NA less than about 0.7. If             worktheobjectiveirisisclosed(just)sufficientlytoexclude
             the microscope is equipped with a phase contrast                        direct rays.
             condenser, the 100 or 40 annuli make excellent,
             centrabledarkfieldpatchstopsforthe(phaseornon-phase)
             4,10 and 20 objectives. If the condenser NA is                       RheinbergIllumination
             greater than 1.0, immersing the condenser to the slide
             (adding one or more drops of immersion oil between the
             condensertoplensandthebottomoftheslide)mayprove                         Replacing the opaque area of the conventional patchstop
             helpful by increasing the effective NA of the condenser.                 by a deep blue transparent filter will change the black
                Since the 1980s, dry darkfield condensers (Figure 3)                  background of a darkfield image into a blue background.
             havebecomeavailable; they are called ‘dry’ to distinguish               Similarly, if a red filter replaces the clear area of the
             them from the immersion systems more fully discussed                    patchstop,thespecimenwillappearred,ratherthanwhite.
             below. The immersion darkfield condensers were quite                     Acolourlessobjectnowappearsbrightredonadeepblue
             difficult to use, mainly because they were so messy; there                background. Beautiful optical staining effects, loved by
             is no such problem with dry systems. In addition, dry                   magazine editors, can be obtained by using combinations
             darkfield condensers do not have the centring pro-                       ofcolouredfiltersinthisway–suggestedfirstbyRheinberg
             blems sometimes associated with the patchstop. These                    in 1896.
             darkfield condensers function by having reflective curved
             surfaces to give highly oblique lighting on the specimen
              Figure 2).
             (                                                                       Practical Applications of Darkfield
             Immersiondarkfieldcondensers(Figure3)                                   Microscopy
             Thesearenotsocommonlyusednowthatdrysystemsare                           At low magnifications, darkfield is extremely helpful to
             commercially available. However, if an oil-immersion                    findandfocusnear-transparentpreparations.Thematerial
             objective with an NA of 1.2 or more must be used for the                doesnotevenhavetobeinfocus,abrightblurshowingits
             observation, then the darkfield condenser must have an                   presence.
             even higher NA, making oil-immersion condensers essen-                     An exciting prospect of darkfield microscopy is that
             tial. These are made with mirrored surfaces so that the                 particles below the resolving power of the objective might
             illuminating cone NAs are approximately 1.2 to 1.4, even                be revealed; it is suggested that the minimum visible
             moreobliquethanwiththedrycondensers.Itisessentialto                     diameterofawhitepointonablackbackgroundislimited
             immersethecondensertotheslide,otherwisetotalinternal                    bytheintensity of illumination, not the NA of the optical
             reflection occurs at the condenser glass/air boundary and                system. More practically, the high contrasts generated by
             no light gets to the slide. Immersed objectives and                     darkfield permit low magnifications, i.e. wider fields, for
             immersed condensers make for troublesome microscopy.                    investigating the presence or absence of very small
             Oil-immersion objectives intended for darkfield work                     pathogens,protozoa,bacteria,etc.(Figure4).Somecaution
             should have an iris diaphragm in the objective bfp. For                 may need to be exercised in interpreting the image.
             brightfieldworktheobjectiveirisisleftopen;fordarkfield                    Although, in the recent past, darkfield has been of prime
             Figure 4   AbloodsmearwithTrypanosomasp.:brightfield(left), darkfield (right). The bar represents 10 mm.
             4
The words contained in this file might help you see if this file matches what you are looking for:

...Light microscopy secondaryarticle brightfield and darkfield article contents kohler illumination theory practice darkfieldillumination ronaldjoldfield macquarieuniversity sydney australia dryandimmersionsystems rheinbergillumination themicroscopeimage especiallyits contrast is profoundly influenced by the way that practical applications of directed on to specimen essential controls required are intensity filters in evenness area angle considerably cost a microscope but extends its performance eciency an system for professional thecondenser mustsatisfy following describes role eld should be fully evenly illumi condenser gives denitions terms used this nated section withhistheoriesofmicroscopelensperformance theintensity illuminated variable ernst abbe determined requirements within wide limits essentially source comple thediameteroftheilluminatedeldshouldbevariable tely ll back focal plane bfp objective match elds view more commonly achieved having each objectives from about mm matching...

no reviews yet
Please Login to review.